HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
Source facts
Action
Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
Initiated
2022-03-14
Root cause
Device Design
Distribution
Worldwide
Recall number
Z-0841-2025
Recall status
Completed
CDRH record ID
211525
FDA classified
2025-01-10
Recall event number
95913
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.