Z-0842-2021: AnshLabs - SARS-CoV2 IgM ELISA (u-Capture), IVD, REF: AL-1002, kit contents, Ab PLATE 1 X ! Plate, IgM Sa

Recall record Z-0842-2021 for Ansh Labs, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0842-2021
Company in FDA record
Ansh Labs, LLC
FDA record creation date
Jan 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling update; To reduce the risk of false positive/false negative results the IFU for the adaptive immune response to SRS-CoV-2 testing is being updated to remove Plasma as a "sample type" and to update the "limitation" section to include potential cross-reactivity with non-SARS-CoV2 strains.

Source facts

Action
On 12/18/2020 and on 12/30/2020, Ansh Labs, sent a "Urgent: Medical Device Voluntary Correction" notification to all affected consignees via E-Mail. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Previous versions of SARS-CoV2 IgM ELISA IFU.AL.1002 (Revision No.: 01 - 04) should be removed and discarded. The attached updated revision, SARS-CoV2 IgM ELISA IFU.AL.1002 (Revision No.: 05) should be used. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 3. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 4. Please complete and return the attached Acknowledgement and Receipt Form within 10 days so we are assured you have received this important communication. 5. If you have any questions, please contact Ansh Labs at 281-404-0260, Monday through Friday 8:30 AM to 5:30 PM, Central Standard Time or at techsupport@anshlabs.com. www.anshlabs.com
Initiated
2020-12-18
Root cause
Device Design
Terminated
2022-01-11
Distribution
US Nationwide distribution including in the states of TX, MO, OK, MD, FL, NY, OR, GA, NC. OUS: None
Recall number
Z-0842-2021
Recall status
Terminated
CDRH record ID
185140
FDA record created
2021-01-11
Recall event number
87057

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗