Open original FDA record ↗Recall record
- FDA identifier
- Z-0843-2012
- Company in FDA record
- Biomet Microfixation, Inc.
- FDA classification date
- Jan 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Biomet Microfixation, Inc., Jacksonville, FL recalled their SternaLock Blu Implant Tray on 11/23/11, model numbers 73-1300 through 73-1302 due to the possibility of an inaccurate reading on the Screw Measuring Area. If used incorrectly, the screw measuring area inside the SternaLock Blu Implant Tray, could lead to the improper selection of a screw 2mm longer than needed.
Source facts
- Action
- Biomet Microfixation, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides a "Recommended Technique" to be followed to correct the problem until replacement trays are available. Contact the Quality Improvement Manager at 1-800-874-7711 for questions regarding this notice.
- Initiated
- 2011-11-23
- Root cause
- Device Design
- Terminated
- 2012-04-03
- Distribution
- Worldwide Distribution-- USA (nationwide) including the states of AL, AZ, CA, CO, DE, FL, GA, HI, IL, IN, KY, MD, MO, MS, NC, NE, NJ, NY, OK, OR, PA, SC, TN, TX, WA, and WI. and countries of Argentina, Australia, Canada, Denmark, Finland, Germany, South Africa, and Spain.
- Recall number
- Z-0843-2012
- Recall status
- Terminated
- CDRH record ID
- 106198
- FDA classified
- 2012-01-24
- Recall event number
- 60682