Z-0843-2020: iChem VELOCITY Urine Chemistry Analyzer

Recall record Z-0843-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0843-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Jan 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity (SG) results.

Source facts

Action
The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letters on 11/07/2019. The letter explained the problem and how to recognize if the S3 flag has been disabled in the instrument. The letter further instructs the following: - Review results to determine if samples report an SG result of >1.060 or an SG result of S3 - If the SG result is >1.060 or S3, test samples for SG using an alternate method* determined by your laboratory - If persistent SG results are >1.060 or persistent S3 results are seen, contact your local Beckman Coulter Representative If you have any questions regarding this notice, please contact Beckman Coulter Customer Support Center: Via our website: http://www.beckmancoulter.com Support > Request Instrument Support. By phone: call 800-526-7694 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter Representative.
Initiated
2019-11-07
Root cause
Error in labeling
Terminated
2023-09-11
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Macau, Malaysia, Maldives, Mexico, Moldova, Montenegro, Morocco, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
Recall number
Z-0843-2020
Recall status
Terminated
CDRH record ID
177875
FDA record created
2020-01-23
Recall event number
84268

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗