Skip to content

Public FDA record

Z-0843-2024: Centered Glenosphere w/Screw, CoCr/TA6V/TiN 10-degrees Tilt 32 mm, REF: 115-3210

Recall record Z-0843-2024 for FX SHOULDER. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0843-2024
Company in FDA record
FX SHOULDER
FDA classification date
Jan 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Centered glenosphere box may contain an eccentric glenosphere.

Source facts

Action
On 8/8/23, the firm telephoned the customer and informed them of the recall. The customer was asked to quarantine the device, which was subsequently collected by the firm.
Initiated
2023-08-08
Root cause
Labeling mix-ups
Distribution
US Nationwide distribution in the state of NJ.
Recall number
Z-0843-2024
Recall status
Open, Classified
CDRH record ID
205322
FDA classified
2024-01-29
Recall event number
93729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗