Z-0844-2019: Vitros Myoglobin Calibrators, Product Code 6801043 For use in the calibration of the VITROS ECi/ECiQ Immu

Recall record Z-0844-2019 for Ortho Clinical Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0844-2019
Company in FDA record
Ortho Clinical Diagnostics Inc
FDA record creation date
Feb 15, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A stability issue in the calibrators may cause calibration failures or atypical Quality Control results following a successful calibration.

Source facts

Action
On November 29, 2018, the firm sent a customer letter all affected customers. All consignees were notified of the issue and informed of revised instructions for VITROS Myoglobin Calibrators: - Reconstitute the calibrators using 1.0 mL distilled water, as per current instructions. - Allow reconstituted calibrator vials to stand at room temperature for 2 hours and then proceed with calibration within 2 hours (4 hours total) - Discard reconstituted calibrators after 4 hours. - Acknowledge the information contained in this notification by completing and submitting the Confirmation of Receipt form no later than December 14, 2018. - Post this notification by each VITROS System at your facility that processes this assay or with the user documentation. - Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Initiated
2018-11-29
Root cause
Under Investigation by firm
Terminated
2024-02-06
Distribution
Distributed nationwide domestically. Foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and the Netherlands.
Recall number
Z-0844-2019
Recall status
Terminated
CDRH record ID
169785
FDA record created
2019-02-15
Recall event number
81767

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗