Open original FDA record ↗Recall record
- FDA identifier
- Z-0844-2024
- Company in FDA record
- Covidien LP
- FDA classification date
- Jan 29, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
Source facts
- Action
- On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact rs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor. Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed. If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1 ***Updated 4/1/2024*** Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024.
- Initiated
- 2023-12-21
- Root cause
- Process control
- Distribution
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
- Recall number
- Z-0844-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205402
- FDA classified
- 2024-01-29
- Recall event number
- 93768