Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Feb 16, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.
Source facts
Action
In the US, beginning 17-Jan-2019, Medtronic Field Representatives (Field Reps) will locate all Encore programmers both in their trunk stock and in consignment locations. Since the Encore Power Supplies are interchangeable among all Encores, all Encore programmer distributed units are being located, and the associated Power Supplies will be checked to identify any affected units. If the Field Rep identifies an affected Power Supply physically at a customer account, the Medtronic Field Rep will hand deliver a courtesy FCA Notification Letter with instructions to the Risk Manager or HealthCare Professional at the account.
Initiated
2019-01-17
Root cause
Device Design
Terminated
2021-03-30
Distribution
US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia,
Syrian Arab Republic
Recall number
Z-0845-2019
Recall status
Terminated
CDRH record ID
170396
FDA record created
2019-02-16
Recall event number
81950