Skip to content

Public FDA record

Z-0845-2019: Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Recall record Z-0845-2019 for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0845-2019
Company in FDA record
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Feb 16, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

Source facts

Action
In the US, beginning 17-Jan-2019, Medtronic Field Representatives (Field Reps) will locate all Encore programmers both in their trunk stock and in consignment locations. Since the Encore Power Supplies are interchangeable among all Encores, all Encore programmer distributed units are being located, and the associated Power Supplies will be checked to identify any affected units. If the Field Rep identifies an affected Power Supply physically at a customer account, the Medtronic Field Rep will hand deliver a courtesy FCA Notification Letter with instructions to the Risk Manager or HealthCare Professional at the account.
Initiated
2019-01-17
Root cause
Device Design
Terminated
2021-03-30
Distribution
US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
Recall number
Z-0845-2019
Recall status
Terminated
CDRH record ID
170396
FDA record created
2019-02-16
Recall event number
81950

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗