Skip to content

Public FDA record

Z-0845-2020: The da Vinci SP surgical system is made of three main components: the Surgeon Console (SC), the Vision Ca

Recall record Z-0845-2020 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0845-2020
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Jan 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm became aware of a problem with the surgical system where the cannula mount button may be damaged when the user releases the button abruptly.

Source facts

Action
On 10/23/2018, the firm sent a "Medical Device Correction" Letter informing them about the cannula mount button issue and informing all personnel who are involved with da Vinci SP Surgery at their site of the following instructions. 1. Release the cannula mount button in a slow and controlled manner. This will prevent the button from snapping back potentially causing button damage. 2. If button damage occurs prior to procedure, discontinue use of the affected SP system. 3. If button damage occurs during a procedure, call dVSTAT for instructions to complete the procedure. ------------ 4. Complete the attached Acknowledgement Form and return it via fax to Intuitive Surgical as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. The firm will: 1. Have a Surgical Representative schedule a site visit to repair your system as soon as possible. 2. Surgical representatives will be available by phone to answer any questions related to this Medical Device Correction. If you need further information or support concerning this Field Safety Notice, customer can contact their Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or ics@intusurg.com
Initiated
2018-10-23
Root cause
Nonconforming Material/Component
Terminated
2022-01-21
Distribution
Product distributed in the US to the following states: Minnesota, New York, Ohio, and Pennsylvania.
Recall number
Z-0845-2020
Recall status
Terminated
CDRH record ID
173971
FDA record created
2020-01-24
Recall event number
83192

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗