The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.
Source facts
Action
The firm implemented site inspections of customers beginning on 11/19/04 which have been completed. Customers that had the suspect component had it replaced.
Initiated
2004-11-18
Root cause
Other
Terminated
2008-01-04
Distribution
Direct accounts located in AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
Recall number
Z-0846-05
Recall status
Terminated
CDRH record ID
38474
FDA classified
2005-05-17
Recall event number
31772