Z-0846-2011: The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified a

Recall record Z-0846-2011 for Philips And Neusoft Medical Systems Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0846-2011
Company in FDA record
Philips And Neusoft Medical Systems Co., Ltd.
FDA classification date
Jan 4, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for the CT patient table top to become detached from the carrier due to missing washers.

Source facts

Action
Neusoft Medical Systems Co., China sent an URGENT DEVICE CORRECTION letter dated Feb. 10, 2010 to their distributor in NC. On 02/12/10, The distributor sent the same letter to their consignees. The firm informed this distributor that there was report of the four screws that secure the patient table top to the carrier that did not have the proper washers in place which caused the table top to detach. If the issue would reoccur, there is a potential for the table top to fall and cause patient injury. Customers were instructed to check the patient table top fixing. If it is found that the the assembly is improper customers were instructed to top using the product and contact local Service personnel for repair. Any customer can contact Service Support Department, Neusoft Medical Systems Co, LTD at nms-service@neusoft.com or Service Support Department at helpdesk@neusoft.com for questions about this recall.
Initiated
2010-02-10
Root cause
Other
Terminated
2011-04-15
Distribution
Nationwide Distribution including NC
Recall number
Z-0846-2011
Recall status
Terminated
CDRH record ID
88867
FDA classified
2011-01-04
Recall event number
54638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗