Skip to content

Public FDA record

Z-0847-2015: Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. N

Recall record Z-0847-2015 for Natus Medical Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0847-2015
Company in FDA record
Natus Medical Incorporated
FDA classification date
Dec 22, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Source facts

Action
Natus Medical has begun mailing the Urgent Recall Notification/Field Safety Notice neoBLUE LED Board Replacements letter (DOC-012737B), dated November 5, 2014, to US consignees. Natus Medical also send the Product Advisory neoBLUE LED Board Replacements letter (DOC-0012738), dated October 30, 2014, to International Distribution Partners. Both US and International consignees also received the TECHNICAL BULLETIN LED Board Kit, neoBLUE 2 LED Phototherapy System, p/n: 001840, Treatment Intensity with revised LED board for neoBLUE 2 systems (p/n: 040904 & 040906) [DOC 008353]; and the Label, neoBLUE Intensity/Height (DOC 011561). Consignees are instructed to measure the light intensity of their devices and to document that information of the device using an adhesive label that is provided. The neoBLUE 3 system and its replacement panel sold today are not affected by this issue. If there are any questions regarding this advisory letter or the attached Technical Bulletin, please call Natus Technical Service at 888-496-2887.
Initiated
2014-11-19
Root cause
Other
Terminated
2015-07-20
Distribution
worldwide
Recall number
Z-0847-2015
Recall status
Terminated
CDRH record ID
131593
FDA classified
2014-12-22
Recall event number
69820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗