Skip to content

Public FDA record

Z-0847-2026: Welch Allyn INF BAG, Thigh 1-Tube, Model/Catalog Number: REF 5089-18; Product replacement bladder for th

Recall record Z-0847-2026 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0847-2026
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Dec 1, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Replacement bladder kit for the Welch Allyn two-piece reusable blood pressure cuff may contain a Large Adult size bladder rather than a Thigh size bladder in the replacement kits.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 11/7/2025 via USPS first class mail. The notice explained the issue, hazard involved and requested the following: For unopened kits: Please inspect any unopened replacement bladder kits you may have. If the product code and lot number (please refer to Figure 1 on page 2 of this letter) on the replacement bladder kit label match those listed in the table above, contact Baxter Technical Support to arrange for return and replacement. For any Thigh size two-piece reusable blood pressure cuffs: Visually confirm the correct dimensions of the bladder; please disregard the lot number printed directly on the bladder. If an incorrect bladder is identified, please contact Baxter Technical Support to arrange for a return and replacement. Contact Baxter Technical Support to arrange for return and replacement product at 800-535-6663 between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday. Press option 2, then select option 1 or 2 for language preference, then press option 2. - If you purchased this product from a distributor or wholesaler, please contact your supplier to arrange for the return and exchange of the affected product. Please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. - Please forward a copy of this communication to the office manager, facility risk manager, patient safety manager, biomedical engineering department, and any other departments within your institution that use the affected product. - If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) and you have affected product please do not distribute. Contact Baxter Technical Support for additional instructions. - If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distribut
Initiated
2025-10-28
Root cause
Process control
Distribution
US Nationwide distribution in the states of California, Colorado, Ohio, and Utah.
Recall number
Z-0847-2026
Recall status
Open, Classified
CDRH record ID
217194
FDA classified
2025-12-01
Recall event number
97926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗