Open original FDA record ↗Recall record
- FDA identifier
- Z-0848-05
- Company in FDA record
- Philips Medical Systems
- FDA classification date
- May 17, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The device may disarm and not deliver shock when performing a Synchronized Cardioversion
Source facts
- Action
- Philips Medical Systems notified accounts by letter on 5/2/05. The firm provided instructions for identifying the effected devices and actions to be taken until the software is availablel for installation.
- Initiated
- 2005-04-29
- Root cause
- Software design
- Terminated
- 2010-09-14
- Distribution
- Nationwide Foreign: Canada, Austria, Australia Bahrain, Belgium, Brazil, China, Hong Kong, India, Finland,France, Germany, Hungry, Israel, Italy, Korea, Kuwait, Latvia, Lebanon, Mexico, Namibia, Libya, Netherland, New Zealnd, Norway, Oma, Philippines, Portugal, Qatat, Russia, South Africa, Singapore Spain, Sweden, Switzerland, Tawian, Thailand, Tunisia, UK, United Arab Emirates, Yemen
- Recall number
- Z-0848-05
- Recall status
- Terminated
- CDRH record ID
- 38809
- FDA classified
- 2005-05-17
- Recall event number
- 31955