Skip to content

Public FDA record

Z-0848-05: Philips Medical HeartStart MRx Monitor/Defibrillator Model: M3535A, M3536A

Recall record Z-0848-05 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0848-05
Company in FDA record
Philips Medical Systems
FDA classification date
May 17, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The device may disarm and not deliver shock when performing a Synchronized Cardioversion

Source facts

Action
Philips Medical Systems notified accounts by letter on 5/2/05. The firm provided instructions for identifying the effected devices and actions to be taken until the software is availablel for installation.
Initiated
2005-04-29
Root cause
Software design
Terminated
2010-09-14
Distribution
Nationwide Foreign: Canada, Austria, Australia Bahrain, Belgium, Brazil, China, Hong Kong, India, Finland,France, Germany, Hungry, Israel, Italy, Korea, Kuwait, Latvia, Lebanon, Mexico, Namibia, Libya, Netherland, New Zealnd, Norway, Oma, Philippines, Portugal, Qatat, Russia, South Africa, Singapore Spain, Sweden, Switzerland, Tawian, Thailand, Tunisia, UK, United Arab Emirates, Yemen
Recall number
Z-0848-05
Recall status
Terminated
CDRH record ID
38809
FDA classified
2005-05-17
Recall event number
31955

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗