Open original FDA record ↗Recall record
- FDA identifier
- Z-0848-2014
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA classification date
- Jan 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Maquet Compressor Mini may deliver compressed air at a temperature higher than specified.
Source facts
- Action
- Maquet sent a product recall urgent-medical device field correction letter/customer fax response via Fed Ex on December 5, 2013, to their customers. Customers are advised of the issues with the device and provided instructions. The correction will include an update with replacement of the affected connector and if needed replacement of the thermoelectric cooler if it damaged. These operations will be performed by an authorized Maquet representative. Any questions can be directed to customer service at 1-888-627-8383 option 2 followed by option 3 Monday through Friday between the hours of 8 am and 6 pm EST. Customers should completed and return the enclosed confirmation form or scan a copy to field actions@maquet.com or fax (973) 807-9290.
- Initiated
- 2013-12-05
- Root cause
- Material/Component Contamination
- Terminated
- 2015-02-05
- Distribution
- Worldwide Distribution US, Canada and Japan.
- Recall number
- Z-0848-2014
- Recall status
- Terminated
- CDRH record ID
- 124699
- FDA classified
- 2014-01-28
- Recall event number
- 67100