Z-0848-2019: VITROS Chemistry Products VALP Reagent Pack. For in vitro diagnostic use only. CATALOG #(s): 6801710

Recall record Z-0848-2019 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0848-2019
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Feb 19, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Negative bias results when using VITROS VALP Reagent, GEN 25.

Source facts

Action
The firm, Ortho Clinical Diagnostics, issued on 11 December 2018, a "URGENT PRODUCT CORRECTION NOTIFICATION" customer letter (Ref. CL2018-267) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who were shipped VITROS Chemistry Products VALP GEN 25. Foreign affiliates were informed by email on 11 December 2018 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. The customers were notified of the issue and instructed to cease use of VALP GEN 25. If you have any GEN 25 (Lot 25-6393) remaining in your inventory, Ortho will provide replacement or credit your account. " Discontinue using Gen 25 (Lot 25-6393), discard any remaining inventory, and switch to GEN 26 or above. " If you do NOT have GEN 26 or above at this time, consider the following alternate options: o Use an alternate valproic acid method until your replacement order arrives. o Consult with your medical director to determine if the bias observed when using GEN 25 is acceptable for continued use, and if so, continue to use until your replacement arrive. In accordance with regulatory requirements, complete and return the Confirmation of Receipt - Response form no later than December 19, 2018 via Fax: 1.888.557.3759 or 1.585.453.4110 or e-Mail Address: RA-OCDUS CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM. " Post this notification by your VITROS System or with your user documentation. " Forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311..
Initiated
2018-12-11
Root cause
Process control
Terminated
2020-09-16
Distribution
Worldwide distribution- US (Nationwide) and countries of: Australia, Belgium, Bermuda, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands and United Kingdom.
Recall number
Z-0848-2019
Recall status
Terminated
CDRH record ID
169917
FDA record created
2019-02-19
Recall event number
81806

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗