Z-0848-2022: TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1

Recall record Z-0848-2022 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0848-2022
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Mar 31, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

Source facts

Action
RaySearch issued Field Safety Notice, Medical Device Correction #83773 notice via email on Sept. 9, 2021. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY THE USER " Avoid changing the dose grid after starting an optimization of TomoHelical and TomoDirect plans with dynamic jaw mode. " Always reset the optimization after editing the dose grid for TomoHelical and TomoDirect plans with dynamic jaw mode. " Always reset the optimization after modifying the target ROI. " It is recommended to use the BEV to inspect the jaw positions superior and inferior to the target dynamic jaw mode. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation, scheduled for market release in October 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact quality@raysearchlabs.com.
Initiated
2021-09-09
Root cause
Under Investigation by firm
Terminated
2024-04-16
Distribution
US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.
Recall number
Z-0848-2022
Recall status
Terminated
CDRH record ID
192432
FDA record created
2022-03-31
Recall event number
89788

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗