Sterilty (package integrity) Compromised: Some failures were for damage to the outer pouch, while another set of failures were for the seals on this pouch. The seal between the inner tray and lid has not been compromised and the contents remain sterile, however, the sterility of the outer surfaces of the inner tray and lid cannot be assured.
Source facts
Action
On 10/4/2007, consignees were notified by Medtronic field personnel, either by personal visits or telephone.
All wholesalers of this product are OUS. Medtronic OUS Regulatory personnel communicated this issue directly to the wholesalers, who provided the letters to their customers. Medtronic field personnel are responsible for completing and returning reply cards confirming all notifications (100%).
Initiated
2007-10-04
Root cause
Package design/selection
Terminated
2012-07-13
Distribution
Worldwide, USA, Taiwan, Thailand, Japan, Spain, Sweden and The Netherlands.
Recall number
Z-0849-2008
Recall status
Terminated
CDRH record ID
66464
FDA classified
2008-02-07
Recall event number
45912
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.