Z-0849-2022: RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14

Recall record Z-0849-2022 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0849-2022
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Mar 31, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated

Source facts

Action
RaySearch issued Field Safety Notice, Medical Device Correction #84236 via E-mail on 22 September 2021. Letter states reason for recall, health risk and action to take: " Use the automatic tool in RayStation/RayPlan to create the External ROI. " When using a Support or Fixation ROI of tissue-like material (density around 1), extend this ROI a few millimeters inside the External ROI to cover the surface voxels of the patient. " If the Support or Fixation ROI differs significantly from the adjacent body tissue, an extra help ROI with material override set to a suitable material, for example, Skin, can be created inside the External ROI. The help ROI must have a width of a few millimeters and extend along the part of the patient surface connected to the Support or Fixation ROI. The same workflow can be used for Bolus ROIs in combination with electron beams. " If the bolus/support/fixation device is part of the CT data, an alternative solution is to include its structure inside the External ROI. In this case the structure should not be defined as a Bolus, Support or Fixation ROI, but as a normal ROI, by setting the ROI type to, for example Other . When included in the External ROI, the user can choose to use the CT densities or to apply a material override to the structure. If this method is applied for a bolus, the user must make sure that only the applicable beams pass through the structure. " For protons and light ions, the user should consider using a higher density uncertainty than normal in robust optimization for active scanning techniques and larger proximal and distal margins for the target in passive techniques, when shooting through a Fixation or Support ROI. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. This issue will be resolved in the next version of
Initiated
2021-09-22
Root cause
Device Design
Distribution
US Nationwide.
Recall number
Z-0849-2022
Recall status
Open, Classified
CDRH record ID
192469
FDA record created
2022-03-31
Recall event number
89800

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗