Open original FDA record ↗Recall record
- FDA identifier
- Z-0850-2010
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Feb 25, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
Source facts
- Action
- An Urgent: Product Corrective Action letter, dated June 9, 2008, was sent to customers. The letter identified the affected products, stated the background, discussed the issue, and stated the actions to be taken. Customers are to: 1. Only use upper case alpha characters in the Sample Identifier 2. If you must use lower case alpha characters in the Sample Identifier, configure the SlideMaker Slide labels without a bar-code. Questions regarding the Product Corrective Action should be directed to Beckman Coulter Customer Service at 1-800-526-7694 or your local Beckman Coulter Rep.
- Initiated
- 2008-05-29
- Root cause
- Pending
- Terminated
- 2011-11-29
- Distribution
- Worldwide Distribution -- USA and Canada.
- Recall number
- Z-0850-2010
- Recall status
- Terminated
- CDRH record ID
- 72444
- FDA classified
- 2010-02-25
- Recall event number
- 48978