Z-0850-2015: Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Ill

Recall record Z-0850-2015 for Illumina Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0850-2015
Company in FDA record
Illumina Inc
FDA classification date
Dec 23, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.

Source facts

Action
A customer notification letter dated 11/12/14 was sent to all customers who purchased the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay. The letter confirms that there is a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. At certain genomic locations within the region sequenced by the assay, deletions cannot be reported by the software. The letter outlines the potential issues, the results of Illuminia's investigation, and the firm's next steps. Customers' with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at techsupport@illumina.com. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at RA@illumina.com. Customers will be contacted by 11/13/14 by phone and a phone script was provided that includes the procedure, talking points, actions required, and next steps.
Initiated
2014-11-06
Root cause
Under Investigation by firm
Terminated
2015-03-17
Distribution
US: MD, NY Non-US: France
Recall number
Z-0850-2015
Recall status
Terminated
CDRH record ID
131410
FDA classified
2014-12-23
Recall event number
69715

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗