Z-0850-2018: Philips Network Firewall (Cisco ASA 5506), Model # 867098

Recall record Z-0850-2018 for Philips Electronics North America Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0850-2018
Company in FDA record
Philips Electronics North America Corporation
FDA classification date
Mar 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.

Source facts

Action
Philips sent an Urgent Recall Notice letter dated October 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that a Philips representatives will contact affected customers to arrange for replacement of all affected units supplied by Philips. For questions contact your local Philips representative: Philips Customer Care Solution Center 800-722-9377.
Initiated
2017-10-05
Root cause
Nonconforming Material/Component
Terminated
2019-07-01
Distribution
US Nationwide
Recall number
Z-0850-2018
Recall status
Terminated
CDRH record ID
160720
FDA classified
2018-03-03
Recall event number
78849

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗