Open original FDA record ↗Recall record
- FDA identifier
- Z-0850-2021
- Company in FDA record
- Penumbra Inc.
- FDA classification date
- Jan 18, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
Source facts
- Action
- On 12/15/2020, the firm posted a press release (https://www.penumbrainc.com/wp-content/uploads/2020/12/JET-7XF-15Dec20.pdf) and issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" letter via FedEx informing customers that all configurations of their reperfusion catheters may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization or contrast injection may result in potential vessel damage, and subsequent patient injury or death. Customers are instructed to: 1. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 2. Immediately review your inventory for the specific catalog numbers listed above. 3. Remove all units from inventory and place them in quarantine for return to Penumbra, Inc. 4. Complete and return the attached product identification / return form within 3 business days. 5. Continue to report adverse events and quality problems experienced with the use of this product. Adverse events and quality problems may be reported to FDA s MedWatch Adverse Event Reporting Program either online, by regular mail, or by fax. For more information, please see https://www.accessdata.fda.gov/scripts/medwatch/index.cfm. The Recalling Firm will contact customers directly to arrange return of affected units and replace/refund returned product. If customers have any questions or concerns, they can contact the Recalling Firm's Customer Service (order@penumbrainc.com or 1.888.272.4606), available Monday - Friday 7:30 AM to 4:00 PM PST, or their sales representative.
- Initiated
- 2020-12-15
- Root cause
- Process design
- Terminated
- 2024-05-14
- Distribution
- Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. And O.U.S (countries) of: Canada, Chile, Colombia, Costa Rica, Dominican Republic, Hong Kong, Israel, Japan, Korea, Lebanon, Malaysia, Nepal, New Zealand, Oman, Panama, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, and Vietnam.
- Recall number
- Z-0850-2021
- Recall status
- Terminated
- CDRH record ID
- 185203
- FDA classified
- 2021-01-18
- Recall event number
- 86989