Possible difficulty in passing the catheter through the needle due to a reduced ID of the needle caused by over crimping of the needle during the manufacturing process.
Source facts
Action
Consignees were notifed by letter on/about 02/17/2005
Initiated
2005-02-17
Root cause
Other
Terminated
2006-05-05
Distribution
Nationwide, Africa , France, Holland, Rotterdam
Recall number
Z-0852-05
Recall status
Terminated
CDRH record ID
37398
FDA classified
2005-06-01
Recall event number
31215