Open original FDA record ↗Recall record
- FDA identifier
- Z-0852-2015
- Company in FDA record
- Getinge Disinfection Ab
- FDA classification date
- Dec 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.
Source facts
- Action
- An Urgent Field Safety Notice was sent to consignees in North America via USPS Certified Mail with signature receipt on 11/25/14 informing them of the correction. The actions that will be carried out will fully address the problem of sedimentation i.e. stop any future sedimentation. You will be contacted by a Getinge representative to schedule a visit. No additional precautions need to be taken by the end user until the Getinge representative arrives. For questions, please contact Getinge USA at 1-800-320-3532.
- Initiated
- 2014-11-26
- Root cause
- Device Design
- Terminated
- 2017-05-12
- Distribution
- Distributed US (nationwide) and the country of Canada.
- Recall number
- Z-0852-2015
- Recall status
- Terminated
- CDRH record ID
- 131600
- FDA classified
- 2014-12-23
- Recall event number
- 69823