Skip to content

Public FDA record

Z-0853-05: Baxter PosiFlow Access Device for IV Access; product code 2N9050; a single use, sterile device with PosiF

Recall record Z-0853-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0853-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Jun 1, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inadequate directions for use resulting in fluid leaks, valves sticking in the open position, and flow output restriction.

Source facts

Action
Urgent Device Correction letters dated 3/3/05 were sent to the direct accounts via first class mail, to the attention of the Director of Nursing. The accounts were informed of the field reports received concerning the PosiFlow Access Device and IV sets containing the PosiFlow device and the changes made to the directions for use and in-service training materials to correct the problems. The accounts were requested to complete and fax back to Baxter at 847-270-5457 the enclosed customer reply form, acknowledging the receipt and dissemination of the information to their staff. Wholesalers were requested to notify their customers. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Initiated
2005-03-03
Root cause
Other
Terminated
2006-05-11
Distribution
Nationwide
Recall number
Z-0853-05
Recall status
Terminated
CDRH record ID
37579
FDA classified
2005-06-01
Recall event number
31275

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗