Incorrect software version of instrument controller is not compatible with new configuration. This incorrect configuration may cause the instrument system to fail at start-up or during a run.
Source facts
Action
Afffymetrix notified the two hospitals by email dated 4/14/05 and by letter to the Distributor on 4/28/05. Affymetrix service personnel will be provided with discs to perform the software update. All field changes will be documented via Service Work Orders.
Initiated
2005-04-26
Root cause
Other
Terminated
2006-12-13
Distribution
California and Indiana
Foreign: The Netherlands, Switzerland
Recall number
Z-0853-06
Recall status
Terminated
CDRH record ID
45455
FDA classified
2006-05-06
Recall event number
35175