Open original FDA record ↗Recall record
- FDA identifier
- Z-0854-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.
Source facts
- Action
- Planned action (s) to repair defect or to bring product into compliance for Siemens Medical Solutions Inc. concerning the corrective action plan (CAP) for the Model(s) Artis Zee/Zeego Stationary Angiographic X-ray Systems. 1.Notifying the customer of the defect. 2.Implementing a corrective software patch to protect the system from damage in the event of hardware failure 3. Scheduling a service visit from a Siemens representative to perform a hardware replacement. 4. Replacing the non-conforming X-ray tube with one manufactured after the nonconformity issue was resolved. For further questions please call (610) 219-6300.
- Initiated
- 2014-09-17
- Root cause
- Process control
- Terminated
- 2015-08-06
- Distribution
- USA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.
- Recall number
- Z-0854-2015
- Recall status
- Terminated
- CDRH record ID
- 130745
- FDA classified
- 2015-01-14
- Recall event number
- 69478