Recall record Z-0854-2022 for Illumina, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0854-2022
Company in FDA record
Illumina, Inc.
FDA record creation date
Mar 31, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

Source facts

Action
On February 22, 2022 Illumina issued a "Urgent Field Safety" recall notice to all affected consignees via E-Mail. The firm issued an amended "Urgent Medical Device" recall notice on March 11, 2022 via email to provide consignees with additional information. In addition to notifying consignees about the recall, the firm asked consignees to take the following actions: 1. Complete the attached FSN2022-125 verification form to confirm receipt of the notification and return the form to Technical Support at techsupport@illumina.com within 5 business days. 2. If you have not already sent Illumina the Verification Form attached to FSN2022-1215, please complete the attached copy of the form and return it to Technical Support at techsupport@illumina.com within 5 business days. No additional action is required at this time 3. Please note, if an operator experiences an adverse event with the use of this product, user may report this to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2022-02-22
Root cause
Device Design
Distribution
US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federation Saudi Arabia Singapore Slovakia South Africa South Korea Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom Vietnam
Recall number
Z-0854-2022
Recall status
Completed
CDRH record ID
192246
FDA record created
2022-03-31
Recall event number
89699

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗