Open original FDA record ↗Recall record
- FDA identifier
- Z-0855-05
- Company in FDA record
- Arizant Inc
- FDA classification date
- Jun 1, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the blanket's air channels, which could have a potential to shift the patient's position on the blanket.
Source facts
- Action
- Recall notification was mailed to customers beginning 03/11/05. Customers were informed of the issue and asked to check their inventory for the affected lots. Affected lots found are to be quarantined and customers are to contact Arizant for instruction on returning affected product and receiving replacement product. Distributors are instructed to examine their inventory for affected lots, discontinue distribution and quarantine the lots. Distributors are asked to contact Arizant for returning affected product. If affected product was distributed, Distributors are to contact Arizant with names, addresses and telephone numbers of their accounts so Arizant can immediately inform those accounts of the recall.
- Initiated
- 2005-03-11
- Root cause
- Other
- Terminated
- 2007-01-10
- Distribution
- nationwide
- Recall number
- Z-0855-05
- Recall status
- Terminated
- CDRH record ID
- 37753
- FDA classified
- 2005-06-01
- Recall event number
- 31379