Open original FDA record ↗Recall record
- FDA identifier
- Z-0855-2007
- Company in FDA record
- Abbott Laboratories MPG
- FDA classification date
- Jun 14, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
False Negative Results : The analytical sensitivity claim of less than or equal to 0.01 ng/mL (ug/L) might not be met for all lots of Architect STAT Troponin-I reagents, thus false negative results may occur with negative results less than 0.1 ng/mL
Source facts
- Action
- Abbott sent a Product Correction letter on 3/8/07 to all Troponin-I customers who received reagent lots in the last 12 months. The letters informed the accounts that the analytical sensitivity claim of = 0.01 ng/mL might not be met for all lots of Architect STAT Troponin-I reagents. The accounts were directed to evaluate the information the information provided in the letter against the reference range used in their laboratory to determine if any action needs to be taken; to follow their facility's policies and procedures for communicating with the heath care providers they serve; to forward a copy of the letter to any laboratory to whom they may have forwarded the reagent; and to retain a copy of the letter for their records. Any questions were directed to Abbott Customer Support at 1-877-4ABBOTT in the U.S. A follow-up press release was issued on 5/25/07 and a follow-up letter was issued on 5/29/07, providing updated instructions for laboratories to follow when using the product.
- Initiated
- 2007-03-08
- Root cause
- Other
- Terminated
- 2008-07-16
- Distribution
- Worldwide: USA including Puerto Rico, and internationally to Canada, Brazil, Germany, Japan, Taiwan, Hong Kong, Thailand, South Korea, Australia, New Zealand, Mexico, Colombia, Chile, Argentina and Trindad & Tobago.
- Recall number
- Z-0855-2007
- Recall status
- Terminated
- CDRH record ID
- 50814
- FDA classified
- 2007-06-14
- Recall event number
- 37570