When using the multiple static segment option, the device may alter dose delivery.
Source facts
Action
An Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps on October 2, 2008. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. The firm recommends that customers DO NOT USE the Multiple Static Segment (MSS) option in Varian Leaf Motion Calculator version 8.2.22. The firm also recommends setting the default delivery method for Varian Leaf Motion Calculator to Sliding Window. When using Eclipse 8.5, always follow the Safety Instructions found in chapter 2 of "Eclipse External Beam Planning Instruction for Use", P/N B501357R01A, March 2008. If customers believe a patient treatment has been affected by this issue, they should notify a Varian Service or Support person as soon as possible.
Initiated
2008-10-02
Root cause
Software design
Terminated
2010-11-27
Distribution
Worldwide Distribution --- including USA and countries of Switzerland and the Netherlands.
Recall number
Z-0855-2009
Recall status
Terminated
CDRH record ID
74812
FDA classified
2009-01-15
Recall event number
50143