Open original FDA record ↗Recall record
- FDA identifier
- Z-0855-2014
- Company in FDA record
- Orthofix, Inc.
- FDA classification date
- Jan 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's collet would malfunction resulting in the surgeon being unable to use the Modular Screw Driver to effectively place Modular Screws, which may result in a delay of surgery.
Source facts
- Action
- Orthofix sent an Urgent Medical Device Recall Notification letter dated November 26, 2013 by certified mail to distributors. The letter identified the affected product, problem and actions to be taken. The letter instructed distributors to immediately cease any further distribution, return affected products to Orthofix at no cost and complete the attached reply form. For questions contact your local Orthofix representative.
- Initiated
- 2013-11-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-12-09
- Distribution
- Worldwide Distribution - US Nationwide including (Puerto Rico) and the countries of: Spain, Germany, Australia, Peru, Mexico, Colombia, South Africa, and Israel.
- Recall number
- Z-0855-2014
- Recall status
- Terminated
- CDRH record ID
- 124351
- FDA classified
- 2014-01-28
- Recall event number
- 66969