Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature
Source facts
Action
BK Medical notified customers by letter on 4/10/06 via FedEx advising user of the problem and scheduling the software correction by a Service representative visit .
Initiated
2006-03-16
Root cause
Other
Terminated
2006-10-12
Distribution
CA,CT, FL, MD, ME, MN, MS, NY, OH, RI, SD, TX, VA
Recall number
Z-0856-06
Recall status
Terminated
CDRH record ID
44908
FDA classified
2006-05-09
Recall event number
34900