Open original FDA record ↗Recall record
- FDA identifier
- Z-0856-2009
- Company in FDA record
- Accuray Inc
- FDA classification date
- Jan 14, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product may not be tensioned properly, potentially causing unexpected rotation or descent.
Source facts
- Action
- All affected sites were notified of the potential problem via Customer Advisory Notice dated 30 October 2008. The letter states required actions that the customer should follow at the beginning of each treatment day. The tension of the belt drive and the support bolts in question will be inspected by Accuray Field Service personnel at the affected facilities. In addition, the preventive maintenance procedure will be updated to include regular inspection and replacement of the belt-drive. Any questions or concerns regarding this issue should be directed to Accuray Customer Support at 1-877-668-8667 or 1-408-716-4700 or customersupport@accuray.com.
- Initiated
- 2008-10-30
- Root cause
- Process control
- Terminated
- 2009-09-10
- Distribution
- Worldwide Distribution --- including USA and countries of France and Turkey.
- Recall number
- Z-0856-2009
- Recall status
- Terminated
- CDRH record ID
- 74987
- FDA classified
- 2009-01-14
- Recall event number
- 50208