This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample processing, the system will post a yellow caution event and continue to operate with the dispense plate in an elevated position until the instrument
Source facts
Action
Beckman Coulter issued an "Urgent: Product Corrective Action" letter dated October 28, 2009. Customers were notified of the affected product and how the issue will be resolved.
For further information, contact Beckman Coulter Inc. Technical Support at 1-800-854-3633.
Initiated
2009-10-28
Root cause
Software design
Terminated
2012-05-07
Distribution
Worldwide Distribution -- United States and Canada.
Recall number
Z-0856-2010
Recall status
Terminated
CDRH record ID
86317
FDA classified
2010-02-25
Recall event number
53679