Open original FDA record ↗Recall record
- FDA identifier
- Z-0858-06
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- May 10, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During use of GE Signa nuclear magnetic resonance imaging system, a flexible Teflon tube that was part of the magnet vent system ruptured and allowed the magnet to vent into the scan room. Because of helium pressure inside the room, the scan room door could not be reopened after the patient was removed from the scan room.
Source facts
- Action
- GE Healthcare is issuing an Urgent Safety Notification to all affected customer sites beginning 09/01/05. The notification describes the concurrent condition that must exist in order for the event to occur. Users are warned to be present at all times when scanning patients and to take emergency measures to vent the scan room should the magent quench. Beginning September 2005, all affected customers'' devices will have the Teflon venting tube replaced with a nonflexible venting tube.
- Initiated
- 2005-08-19
- Root cause
- Other
- Terminated
- 2007-02-03
- Distribution
- Nationwide and to countries OUS including: United Arab Emirates, Argentina, Australia, Brazil, Canada, China, Germany, Denmark, Egypt, Spain, France, United Kingdom, Ireland, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Saudi Arabia, Sweden, Turkey and Tiawan
- Recall number
- Z-0858-06
- Recall status
- Terminated
- CDRH record ID
- 41690
- FDA classified
- 2006-05-10
- Recall event number
- 33320