Z-0858-2022: 1.5T SIGNA Excite HD, 3.0T SIGNA Excite HD. Magnetic Resonance Imaging System

Recall record Z-0858-2022 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0858-2022
Company in FDA record
GE Healthcare, LLC
FDA classification date
Apr 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for the images to be flipped left to right.

Source facts

Action
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Initiated
2022-02-18
Root cause
Device Design
Distribution
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-0858-2022
Recall status
Open, Classified
CDRH record ID
192303
FDA classified
2022-04-04
Recall event number
89720

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗