Skip to content

Public FDA record

Z-0859-06: Cool-tip RF Generator, for use in percutaneous, laparoscopic, intraoperative coagulation and ablation of

Recall record Z-0859-06 for Covidien LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0859-06
Company in FDA record
Covidien LP
FDA classification date
May 10, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Electrosurgical ablation device may experience loss of power during use, when used with Cool-tip Switching controller in ablation mode.

Source facts

Action
Consignees were notified by telephone followed by a fax on 04/21/2006.
Initiated
2006-04-21
Root cause
Other
Terminated
2009-02-26
Distribution
Nationwide. AZ, CA, CO, CT, FL, GA, IL, MA, MI, MN, MO, MS, NC, NE, NJ, OH, OK, PA, TX, VA. Foreign distribution to Argentina, Canada, Chile, Costa Rica, France, Greece, Israel, Peru, Singapore.
Recall number
Z-0859-06
Recall status
Terminated
CDRH record ID
45717
FDA classified
2006-05-10
Recall event number
35283

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗