Open original FDA record ↗Recall record
- FDA identifier
- Z-0859-2008
- Company in FDA record
- Zimmer Orthopaedic Surgical Products
- FDA classification date
- Mar 6, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sterility (package integrity) compromised; The effect of a silicone stain produced during assembly operations on the sterility barrier properties of the Tyvek lids has not been validated by the firm.
Source facts
- Action
- Zimmer notified their customers via an Urgent Device Recall letter dated January 8, 2008. The letter requests each customers to inventory their stock. If any recalled product is located the customer is to notify Zimmer via fax and a replacement part will be shipped to them.
- Initiated
- 2007-12-18
- Root cause
- Other
- Terminated
- 2012-05-03
- Distribution
- Worldwide-USA, forty-two international and one Canadian account, including countries of Singapore, Hong Kong, Lebanon, Finland, Australia, Japan, Israel, Korea, Saudi Arabia, Dubai, Denmark, Czech Republic, Guatemala, Brazil, Taiwan, China, Nicaragua, Virgin Island, Switzerland, Austria, Spain, Belgium, The Netherlands, France, Germany, Portugal, England, Sweden, India, Argentina, Italy, Chile, Thailand & Saudi Arabia
- Recall number
- Z-0859-2008
- Recall status
- Terminated
- CDRH record ID
- 66970
- FDA classified
- 2008-03-06
- Recall event number
- 46206