Z-0859-2021: icumedical Cogent" Hemodynamic Monitoring System - Product Usage: intended for patients for whom the moni

Recall record Z-0859-2021 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0859-2021
Company in FDA record
ICU Medical, Inc.
FDA record creation date
Jan 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a potential software issue, the display may show the incorrect continuous cardiac output (CCO) values after PulseCO calibration.

Source facts

Action
On 11/09/2020, the firm's sent an "URGENT: MEDICAL DEVICE CORRECTION" Notice via traceable mailing informing them that the Recalling Firm has identified a software issue with it Hemodynamic Monitoring System which could potentially display the incorrect continuous cardiac output (COO) values after calibration. The Recalling Firm is informing it customer that if the following calibration steps are taken: Step 1: Start PulseCO monitoring by pressing the CCO button which forces a zeroing of the arterial pressure waveform; press zero on Cogent then reinitiate arterial pressure monitoring. Step 2: Wait for the PulseCO cardiac output value to stabilize. Step 3: If the displayed PulseCO value is different than expected, apply a desired calibration value that is different from the nominal PulseCO value. Step 4: The calibration value will be displayed in 10-20 seconds or less from the time of entry. Two potential situations can occur in relation to the calibration process: Situation 1: If PulseCO monitoring is stopped after Step 3 and before the PulseCO value reaches the desired calibration-value in Step 4, followed by pressing the CCO button to restart PulseCO monitoring, in some cases the displayed PulseCO will not reach the desired calibration-value previously entered. The displayed CCO value will be neither the original value nor the desired calibrated value; but rather it will be a value between the two. Situation 2: After completion of Steps 1 4, if PulseCO monitoring is stopped and then an OpMod with a different patient ID stored on it is connected, the Cogent will display the Patient Selection screen. On the Patient Selection screen, the user selects Current Patient and then presses the CCO button to re-start the PulseCO monitoring. In some cases, the displayed PulseCO will not reach the desired calibration value previously entered and could be higher or lower than the entered value. The required actions for Users: There is no need to discontuine use
Initiated
2020-11-09
Root cause
Software design
Terminated
2023-04-19
Distribution
U.S. Nationwide distribution including in the states of AL, AR, CA, CO, FL, IL, KY, MD, ME, MI, NC, NJ, OH, PA, TN, TX, VT, WA and WI. O.U.S.: None
Recall number
Z-0859-2021
Recall status
Terminated
CDRH record ID
184611
FDA record created
2021-01-14
Recall event number
86816

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗