Fixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type.
Source facts
Action
Siemens Healthcare sent a "SAFETY ADVISORY NOTICE" dated December 12, 2011 to the affected customer. The letter identifies the product, problem, and actions to be taken by the customer. The firm prepared a modification, which will be provided to the affected customer under Update Instruction XP055/11/S. The firm will resolve this potential issue by preventively exchanging all bolts for the fixation of the X-ray tube holder of the concerned AXIOM Aristos FX Plus unit.
Initiated
2011-12-12
Root cause
Process control
Terminated
2014-01-14
Distribution
Product distributed in one hospital in California.
Recall number
Z-0860-2012
Recall status
Terminated
CDRH record ID
106477
FDA classified
2012-01-25
Recall event number
60789