A specific subset of model 7428 Kinetra implantable neurostimulators may fail due to lifted wirebonds between the hybrid circuit and battery. This failure mechanism may present clinically as loss of functionality and a return of the patient's underlying medical conditions.
Source facts
Action
An Urgent Device Recall Important Patient Management Information letter was sent to implanting and following physicians of patients who have received the device. The letter describes the issue and provides recommendations.
Initiated
2006-04-04
Root cause
Other
Terminated
2006-12-26
Distribution
USA--States of CA, OR, PA. Internationally -Belgium, France, Germany, Italy, Spain, Switzerland, United Kingdom
Recall number
Z-0861-06
Recall status
Terminated
CDRH record ID
45370
FDA classified
2006-05-27
Recall event number
35110