Z-0861-2024: Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

Recall record Z-0861-2024 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0861-2024
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Jan 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.

Source facts

Action
The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 12/08/2023 by certified mail to consignees. Consignees with affected devices are asked to utilize the provided workaround instructions. Additionally, consignees are asked to make all users aware of this recall notification and keep a copy of this notification with system documentation. Getinge Field Service Engineers will be dispatched to the field to install software updates to affected devices. Consignees with any questions can contact the recalling firm by phone at 1-800-722-9377. Consignees are asked to return the provided response form to the recalling firm at ITG_Recalls@philips.com.
Initiated
2023-12-08
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
Recall number
Z-0861-2024
Recall status
Open, Classified
CDRH record ID
205331
FDA classified
2024-01-30
Recall event number
93734

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗