Z-0862-06: REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Contr

Recall record Z-0862-06 for Konica Minolta Medical Imaging USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0862-06
Company in FDA record
Konica Minolta Medical Imaging USA, Inc.
FDA classification date
May 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will populate the fields where no data has been entered with data from the previous patient.

Source facts

Action
The firm phoned the customers on 2/6/2006 to inform them of the recall and the corrective action associated to the recall. This will be followed up with a fax, e-mail and letter. The final step will be visiting the location to install the correct software.
Initiated
2006-02-06
Root cause
Other
Terminated
2008-08-22
Distribution
The units were distributed to hospitals nationwide through the firm''s sales force and resellers. There are no government accounts.
Recall number
Z-0862-06
Recall status
Terminated
CDRH record ID
44641
FDA classified
2006-05-16
Recall event number
34785

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗