If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will populate the fields where no data has been entered with data from the previous patient.
Source facts
Action
The firm phoned the customers on 2/6/2006 to inform them of the recall and the corrective action associated to the recall. This will be followed up with a fax, e-mail and letter. The final step will be visiting the location to install the correct software.
Initiated
2006-02-06
Root cause
Other
Terminated
2008-08-22
Distribution
The units were distributed to hospitals nationwide through the firm''s sales force and resellers. There are no government accounts.
Recall number
Z-0862-06
Recall status
Terminated
CDRH record ID
44641
FDA classified
2006-05-16
Recall event number
34785