Skip to content

Public FDA record

Z-0862-2010: Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, REF 011901, Sterile EO, Assembled In Mexic

Recall record Z-0862-2010 for Bard Access Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0862-2010
Company in FDA record
Bard Access Systems
FDA classification date
Feb 26, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.

Source facts

Action
Bard Access Systems, Inc. issued and "Urgent: Huber Plus Non-Coring Needle Safety Infusion Set Recall" notification beginning January 2010. Users were instructed to quarantine all affected product and to notify any of their customers that had received the product. Quarantined product is to be returned to the firm. For further information, contact Bard Access Systems, Inc. at 1-800-290-1689.
Initiated
2010-01-14
Root cause
Component change control
Terminated
2010-05-14
Distribution
Worldwide Distribution -- United States (VA, KY, MA, NJ and TN) and Belgium.
Recall number
Z-0862-2010
Recall status
Terminated
CDRH record ID
88628
FDA classified
2010-02-26
Recall event number
53341

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗