Product was packaged with a four inch needle in stead of the correct five inch needle. There is a potential that the needle may not be able to reach the fracture site, which may require additional medical intervention. In addition, if the stylet needs to be changed and the additional stylet that comes in the package is used, this additional stylet is one inch longer than the needle. which could l
Source facts
Action
Stryker sent an Urgent Recall Notification letter dated December 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. If product was further distributed, customers should forward the recall notification letter and the attached Business Reply Form (BRF) to all affected locations. Immediately check inventory and quarantine any affected product found. Mark product as "RECALLED PRODUCT". Indicate on the BRF the quantity of affected PCD kits they are returning and fill in the form completely.
Fax the Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762. Upon receipt of the BRF, Stryker will contact them to coordinate the return of all affected product that they have on hand.
For questions regarding this recall call 269-389-4354.
.
Initiated
2011-11-30
Root cause
Component change control
Terminated
2012-08-14
Distribution
Nationwide Distribution including FL, PA, NY, OH, IL, KY, MT, WA, CA, MO, HI, IN, SC, MS, WV, KS, TX, and WY
Recall number
Z-0862-2012
Recall status
Terminated
CDRH record ID
105884
FDA classified
2012-01-25
Recall event number
60586