Z-0863-06: CONTAK RENEWAL 3 RF (models H210, H215), CONTAK RENEWAL 3 RF HE (models H217, H219), CONTAK RENEWAL 4 RF

Recall record Z-0863-06 for Guidant Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0863-06
Company in FDA record
Guidant Corporation
FDA classification date
May 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Guidant received 39 reports of devices that exhibited lower than expected battery voltage prior to implant; none of these devices were implanted.

Source facts

Action
Affected physicians were sent an Urgent Medical Device Safety Information & Corrective Action letter, dated 05/11/06. A press release was issued March 13, 2006 The letter describes the issue and emphasizes the importance of checking battery voltage prior to implant as stated in device labeling. For implanted devices, continue normal follow-up.
Initiated
2006-03-11
Root cause
Other
Terminated
2006-12-31
Distribution
Nationwide, Puerto Rico, Virgin Islands and to countries that include: Austria, Belgium, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Iran, Ireland, Israel, Italy, Luxembourg, Mexico, Netherlands, Norway, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom
Recall number
Z-0863-06
Recall status
Terminated
CDRH record ID
44837
FDA classified
2006-05-16
Recall event number
34860

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗