Z-0863-2009: DeVilbiss Model 7310PR-D VacuAide Compact Suction Unit. A portable, AC-powered or compressed air-powered

Recall record Z-0863-2009 for Sunrise Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0863-2009
Company in FDA record
Sunrise Medical
FDA classification date
Feb 10, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Improper suction: Regulator knob does not turn properly which can affect the suction level.

Source facts

Action
DeVilbiss notified dealers and users (through the dealers) to recall the units, by a telephone call October 27, 2008 followed by an Urgent Medical Device Recall mailing November 3, 2008. The letter advised customers what models are affected and how to contact the firm of dealer to have the affected unit replaced. Corrective and preventative action on current manufacturing has been implemented.
Initiated
2008-10-27
Root cause
Device Design
Terminated
2012-09-25
Distribution
Worldwide Distribution: USA, Mexico, Guatemala, Colombia, and Costa Rica.
Recall number
Z-0863-2009
Recall status
Terminated
CDRH record ID
75692
FDA classified
2009-02-10
Recall event number
50522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗