Open original FDA record ↗Recall record
- FDA identifier
- Z-0863-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Mar 4, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Excessive Buildup of protein, bacteria and sample tube additives in the ISE flow cell may cause erroneous NA (sodium) results.
Source facts
- Action
- An Urgent: Product Corrective Action letter dated January 4, 2010, was mailed to customers advising of erroneous sodium results derived from build up of protein, bacteria and additives in the ISE flow cell. The letter discussed the impact and actions to be taken along with some additional recommendations. Customers are advised to follow the new maintenance instructions included with the letter. The information in the letter should be shared with laboratory staff and the notification should be retained as part of the Quality System documentation. A copy of the letter should be provided to any other laboratory that affected product was forwarded to. The enclosed response form should be completed and returned. Questions should be directed to the Customer Support Center at 1-800-854-3633.
- Initiated
- 2010-01-04
- Root cause
- Device Design
- Terminated
- 2012-04-17
- Distribution
- Worldwide Distribution -- USA and Canada.
- Recall number
- Z-0863-2010
- Recall status
- Terminated
- CDRH record ID
- 88235
- FDA classified
- 2010-03-04
- Recall event number
- 54344